Pentax Duodenoscope Flaw Linked to Infection Risk

hospital patient's hand over the hospital bed

Pentax reported its ED-3490TK duodenoscope had additional design flaws that could lead to patient infection, according to a U.S. Food and Drug Administration Safety Communication.

In February 2016, the FDA validated Pextax’s updated duodenoscope cleaning instructions. The instruction update came in response to the FDA’s February 2015 Safety Communication linking the scopes to bacterial infections. The design of many endoscopic retrograde cholangiopancreatographic (ERCP) endoscopes may cause bacteria to remain where device parts connect, even after cleaning.

Sometime after February 2016, Pentax told the FDA that “cracks and gaps in the adhesive that seals the device’s distal cap to its distal tip can occur, which can lead to microbial and fluid ingress.”

The company also issued a Field Correction to its customers advising them of the adhesive problem. It urged customers to stop using any devices which have cracks, holes, scratches or other defects because of potential device contamination.

“Pentax will provide annual inspection and servicing to all customers with ED-3490TK duodenoscopes,” the company said in the Field Correction. “During on-site inspections, Pentax field staff will assess duodenoscope functionality and will send duodenoscopes for in-house servicing to recondition worn parts, as needed.”

The FDA said it will continue to monitor the situation and urged health care providers to properly follow all of Pentax’s cleaning instructions.

Contaminated Scopes Lead to Deadly Infection

Health care providers perform about 700,000 ERCP procedures using scopes each year.

Doctors use the tiny cameras attached to the scopes to perform minimally invasive surgeries and diagnostics. They can remove gallstones, treat pancreatitis and drain bile ducts.

After a superbug outbreak in January 2015 at UCLA Medical Center killed several patients and infected about 200 more, the FDA issued a Safety Communication advising health care providers of the problems with contaminated scopes.

Olympus recalled thousands of its TJF-Q180V scopes linked to the outbreak. Infected patients filed superbug lawsuits against Olympus which manufactures the majority of duodenoscopes in the U.S.

Fujifilm and Pentax also faced scrutiny from the FDA.

Fujifilm Pulls Duodenoscopes from Market

Right before Pentax alerted the FDA to additional problems with its duodenoscope, Fujifilm announced it was removing four of its models from the market.

The Japanese company said they would remove the following older models: Fuji ED-250XL5, ED-250XT5, ED-450XL5 and ED-450XT5. In addition, the company said it would replace these models with the newer ED-530XT along with accessories at no cost.

“After discussions with the U.S. Food and Drug Administration (“FDA”), FMSU-ESD has made a business decision to replace these older duodenoscope models with the currently marketed model, the ED-530XT,” Fujifilm said in its letter.

The company stressed that it did not receive adverse events reports and its decision to remove the devices is not because of a “known safety risk.”

Last modified: February 21, 2017

Author

Michelle Y. Llamas is a senior content writer. She is also the host of Drugwatch Podcast where she interviews medical experts as well as patients affected by drugs and medical devices. She has written medical and legal content for several years — including an article in The Journal of Palliative Medicine and an academic book review for Nova Science Publishers. With Drugwatch, she has developed relationships with legal and medical professionals as well as with several patients and support groups. Prior to writing for Drugwatch, she spent several years as a legal assistant for a personal injury law firm in Orlando. She obtained her English – Technical Communication degree from the University of Central Florida. She is a committee member with the American Medical Writers Association.

Hide Sources

  1. Fujifilm. (2017, January 13). Medical Device Removal ED-250XL5/XT5 and ED-450XL5/XT5 Duodenoscopes. Retrieved from https://www.fujifilmusa.com/products/medical/endoscopy/pdf/ED250_and_450_Duodenoscopes_Removal_011217.pdf
  2. Healio. (2017, January 17). Pentax alerts FDA to additional scope design flaw linked to infection risk. Retrieved from http://www.healio.com/gastroenterology/interventional-endoscopy/news/online/%7B43cf9c91-9605-447e-94fe-82fb7dae5019%7D/pentax-alerts-fda-to-additional-scope-design-flaw-linked-to-infection-risk
  3. FDA. (2017, January 17). UPDATE: Importance of Following Validated Reprocessing Instructions for PENTAX ED-3490TK Video Duodenoscopes: FDA Safety Communication. Retrieved from http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm537092.htm
  4. Pentax. (2017, January 16). FIELD CORRECTION Importance of following Validated Reprocessing Instructions PENTAX Medical ED-3490TK Duodenoscope. Retrieved from https://pentaxmedical.com/pentax/en/99/1/FIELD-CORRECTION-Importance-of-following-Validated-Reprocessing-Instructions-PENTAX-Medical-ED-3490TK-Duodenoscope-
Free Superbug Case Review

Were you infected by a superbug after an ERCP duodenoscope procedure?
Free Case Review